Choose the wedge materiel used for the energy evaluation.
The
Density (g/cm3), CSDA r0 g/cm2)and Angle (degrees) fields are automatically completed. Density and CSDA r0 values are fixed but angle value can be changed.
When the heading is complete, the software checks the communication between the PC, the spectrophotometer, and the motor.
Make sure that the motor is installed in the spectrometer. If not, install it. Make sure that for absorbance, the spectrometer is correctly zeroed without dosimeter but with the motor.
If, while the measurements the communication should be interrupted (RS232, Power supply ...) the Aer'ODE Software will detect the problem, save the data and advise the user.
The current wavelength and absorbance appear on the spectrophotometers screen.
A new sheet is displayed on the screen :
- Input the strip dosimeter in the motor. (As written on the screen)
- Click the measure button to start measurement.
The dosimeter strip should not be shorter than 15 cm, for this readout equipment. If the dosimeter strip stacks in the drive motor, then stop the measurement (
Stop button), take out the strip (
Motor Rewind button), cut properly the end of it and try again by a click on the
measure button.
During measurement, the absorbance values are displayed on this window.
The end of the measurement is detected automatically when the absorbance value is near zero or stable for a period of time. It can be also decided by the user with a click on the
stop button.
- At the end of measurement, close the window (as written on the screen) by a click on the close button.
The specific absorbance, absorbed dose and uncertainty is automatically calculated and displayed in the table. (See
annexe about uncertainty calculation).
The Min, Max and mean dose are displayed in the corresponding fields.
- Click the Graphic button to display the graphic.
- Click the energy button to calculate the energy of the beam. The corresponding fields are complete.
In order to make a precise energy measurement, the depth zero position must be set correctly. If a mark is made on the strip before irradiation, a dose peak will appear on the screen. Position this mark at depth zero then it can be used as the position of the entrance of the strip inside the wedge. Care should be taken also in comparing the real length of the dosimeter and the measured length. This can be easily done by making a second mark on the strip before irradiation and measuring by difference the length between the two marks. Should this comparison be out of a specific range (
±2%) then the
step length has to be updated in the
parameter setting routine (See section
Parameter Setting for more informations).
Ep (MeV) and Ea (MeV) fields are completed in the energy determination sheet.