When main menu is displayed and "Instrument Check" selected, choose "Thickness check" on the screen to have access to this sheet.
Complete the heading (mode and operator). The thickness gauge serial number is automatically completed with the informations saved in the registration routine.
Tolerance limits can be modified by a click on the show button. A new table is displayed on the screen
* Select with the mouse the Ref ID to modify and click the

button. Enter the updated value with the keyboard.
* Clear the old value (click the
clear button) display on the measurement sheet and activate the new one with the
load button (which has take the place of the
clear button).
* Close the ref. tolerances table (click the
close button).
* Type
TAB key to progress in the sheet.
The software checks the communication between the PC and the thickness gauge.
Put the first reference thickness block under the thickness gauge (as written on the screen) and type TAB key.
* The thickness value is automatically displayed in the thickness cell for the first measurement. Repeat the measurement with the same block as written on the screen.
After the fifth measurement, the mean value is calculated and the conclusion is automatically completed (OK or not OK).
* Repeat the measurement with the available blocks as written on the screen.
At the end of the check, enter the quality assurance manager name and save the data.